Eye Testing for Kidney Function Rate


  1. Prepare the eye images before inserting them into the KFR. (refer capturing eye images guidelines) Both the left and right eyes should be identified correctly. The following image shows a typical left eye (Fig. 1a) and right eye (Fig. 1b)

  2. figure 1

    (a)                             (b)
    Figure 1
  3. The main user interface can be seen below. (Fig. 2)
  4. figure 2

    Figure 2
  5. Browse for Left eye from your PC using the "Browse Left Eye" button. AA pop-up file explorer window will be opened. . Choose the desired Left eye image and press the open button on the lower right side. (Fig. 3)
  6. figure 3

    Figure 3
  7. Follow the step 3 for the Right eye selection also using "Browse Right Eye" button and add it to the main user interface. The user interface will be shown as below after adding both the images. (Fig. 4)
  8. figure 4

    Figure 4
  9. Use "Crop Image" buttons to automatically detect the Iris and crop the required region of interest. (Fig. 5)
  10. figure 5

    Figure 5


    figure 6

    Figure 6

    If the image does NOT contain sclera or skin you can proceed.


  11. Press "Run" button in order to get the results for the selected subjects' eyes. The Results can be seen in the "Results" section, and a complete log along with other useful information can be found in the "Log" section. (Fig. 7) Refer "Results and Log Values Guideline" for a detailed explanation of the log results.
  12. figure 7

    Figure 7

Copyright © 2017-2020 Eye Tech LK. All rights reserved